Specifications
Dead apace:
Non‐rebreathing valve
7ml
Adult mask
150ml
Child mask
9 5 m l
Infant mask
2 8 m l
Pressure Relief
4 0 ± 5 c m H
60±1 0 c m a H
Reservoir volume
2 0 0 0 m l (adult model)
1 0 0 0 m l (c hi l d model)
5 0 0 m l (Infant model)
Bag volume/Stroke volume
1 5 0 0 / 1 3 5 0 m l ( a d u l t m o d e l )
6 0 0 /
2 8 0 /
Maximum BPM
4 5 b r e a t h s / m i n ( a d u l t m o d e l )
1 0 5 b r e a t h s / m i n ( c h i l d m o d e l )
9 8 b r e a t h s / m i n ( I n f a n t m o d e l )
Oxygen Concentration
With reservoir
9 9 %
Without reservoir
4 5 % ( a d u l t a n d c h i l d m o d e l )
9 0 % ( i n f a n t m o d e l )
0 (infant & child)
2
0 (adult)
2
3 5 0 m l ( c h i l d m o d e l )
1 6 0 m l ( i n f a n t m o d e l )
Testing The Resuscitator
The MR‐100™ Plus Manual Resuscitator should be tested as follows:
when first using the new Resuscitator
after cleaning and disinfection
after any new parts having been fitted
monthly, if the Resuscitator is not frequently used.
Equipment required: Test lung, 0‐100cmH
Child resuscitators only) , flow meter, regulated gas supply & gas supply
tubing.
Testing the silicone bag assembly
1. Remove the non‐rebreathing valve and the oxygen reservoir (if fitted).
2. Compress the silicone bag and occlude (block) the non‐rebreathing valve
outlet.
3. Release the bag. The bag should expand immediately and refill. If not,
check the intake valve at the base of the silicone bag is correctly
assembled or not.
4. While keeping the non‐rebreathing valve outlet blocked, compress the
bag again. The bag should not be compressed easily If this occurs, check
that you are blocking the valve sufficiently, and that the intake valve at
the base of the silicone bag is correctly assembled.
Non‐rebreathing valve assembly
1. Connect the non‐rebreathing valve to the silicone bag. Connect the test
lung to the outlet on the non‐rebreathing valve and is correctly
assembled.
0 manometer (for Infant and
2