Connect the AC POWER CORD to supply. GROUND conductor MUST BE
Used.
Connect Handpiece leads, and Power Control pot leads according to
drawing FRONT VIEW. There is NO POLARITY for Handpiece leads or
Power Control Pot leads. IT IS SUGGESTED THE 15' EXTENSION WIRES
(not supplied) CONNECTED TO THE AUTOSCALER® CONSOLE BE
SOLDERED TOGETHER, AND INSULATED WITH A "UL or CSA"
APPROVED SHRINK TUBING Hand twisting and taping of wire
connections is NOT recommended. It may affect the performance and
safety of the installation.
When installing the AUTOSCALER® build-in unit, the AIR SOURCE used
to activate the scaler MUST be independent of all other operatory
equipment, and the supply line pressure must relieve to ZERO PSI after
each activation. The AUTOSCALER® unit may be damaged by activation
while other equipment is in use if the air signal is not turned OFF to
the scaler, and line pressure fully relieved when not in use.
The AUTOSCALER® is internally fitted from the factory with a pressure
relief valve (restrictor) and must not be removed. This restrictor helps
relieve line pressure after each activation.
IMPORTANT: The AUTOSCALER® unit has been calibrated
anticipating approximately 15' Fifteen feet of 18 GAUGE wire will be
used in the installation. DO NOT use any wire thinner than 18 GAUGE.
This will change the tuning (i.e. insert will not vibrate properly) of the
AUTOSCALER® and/or not function.
25

ULTRASONIC SCALER INSERTS

DIRECTIONS FOR USE FOR OPTIMAL FUNCTION.
Sterilize before each use in accordance with the insert
manufacturer's instructions
Federal Law restricts the device to sale by or on the order of a
licensed practitioner
3 month warranty
INSERT DEVICE DESCRIPTION
Ultrasonic inserts are accessory inserts designed for use with
magnetostrictive scaler units, (see specifications) that generate a
frequency of either 25KHz or 30KHz. A range of tip designs are available to
suit different clinical conditions. Finger-grips are either plastic or metal
and the water channel is either of internal or external design. These
inserts are designed to be used in magnetostrictive ultrasonic scalers only
by professionally trained operators, such as Dentists or Dental Hygienists,
to perform general supra and subgingival scaling procedures.
Internal water channel with plastic finger-grip
Tip Design
DF#3 Beaver Tail/Flat
Heavy to moderate supragingival deposits
on buccal and lingual surfaces.
DF#10 Universal Curette
Heavy to moderate supragingival deposits,
light to moderate subgingival deposit on all
surfaces.
DF#100 Periodontal Style
Light to moderate supragingival deposits
especially interproximally.
External water channel with metal finger-grip
Tip Design
DF#10 Universal Curette
Heavy to moderate supragingival deposits,
light to moderate subgingival deposit on all
surfaces.
DF#100 Periodontal Style
Light to moderate supragingival deposits
especially interproximally.
Application
Application
26