Pair
Step 1
Install myFreescan app
on your smartphone
Install myFreescan app on your smartphone.
Scan the below QR code or search on App store / Google Play for
myFreescan app to install it on your smartphone.
Download for
Apple iOS
Step 2
Pair device and
myFreescan app
Eva
3
Ensure your smartphone Bluetooth is enabled.
– Click "Setting" on the dashboard of myFreescan app.
– Click "Bluetooth setting".
– Click the "Pair" button and search for a pairable
device.
01-01 05:23
PM
A
1234-5678
USER A
Accept
BT
Reject
Turn on the power of
Pick your
Freescan
the device and press
device and verify the
the down button "V"
passkey to get paired.
to enable Bluetooth.
Step 3
Upload the measured data
to myFreescan app
Please ensure your device is
turned on and click "Sync" to
upload the measured data.
D
Calibrate
can be calibrated with another certified
blood pressure monitor to ensure its accuracy.
Please calibrate your
the first time, after dramatic weight change, surgeries, or
critical illness, and calibrate it every 6 months afterwards.
Round 1
01-01 05:23
PM
A
USER A
A
4
Calibration
Round 1/2
Start
measuring
with Freescan
Calibration
Measure with
Certified BP
Monitor.
Done
Adjust to
Round 1 result
Round 2
Calibration
Round 2/2
Start
measuring
with Freescan
Calibration
Measure with
Certified BP
Monitor.
Done
Adjust to
Round 2 result
E
rror message
E01
1. Rest your shoulder, forearm and
elbow comfortably on the table .
2. Make sure your hands are not
off the table.
3. Do not shake the device.Do not
Device not
attempt to view the device screen.
stable
Press the Start button to measure
again.
before using it for
Keep the Start
until the measurement is complete.
Press the Start button to measure
Start button
again.
released
Calibration
Set
Clear
Enable the Bluetooth function on
your smartphone.
Ensure the distance of device and
Back
A
smartphone are within 5 meters.
Bluetooth
not found
Press the Start button to measure
again.
1. Use wet wipes to moisturize the
wrist and nger tips.
2. Wipe away grease on the
electrodes.
3. Con rm Step 3 and Step 4 of B
No ECG
detected
Press the Start button to measure
again.
E05
Con rm Step 2 of the location of
your pulse and con rm Step 4 of
A and B.
Weak pulse
Press the Start button to measure
again.
If E05 still appears, rest for 5 minutes and try to measure again.
Please reset the device through
the reset hole. If this error reoccurs,
please contact the distributor for
further assistance.
Device
failure
Press the Start button to measure
again.
E07
Keep at least one meter away
from radio sources, such as
computer screens.
Radio
Press the Start button to measure
interference
again.
If E07 occurs constantly, please contact the distributor for further
assistance.
User Switching
01-01 05:23
01-01 05:23
PM
A
USER A
button pressed
Press and hold the up button to switch to the
other user if created.
FCC Notices :
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.This equipment has been
tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio frequency energy and, if
not installed and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television reception, which
can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures :
Reorient or relocate the receiving antenna.Receiver is connected.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
CAUTION :
Any changes or modifications not expressly approved by the grantee of this device could void the user's
authority to operate the equipment.
Model: Freescan
Product Name: Heart Performance Monitor (311020)