V S A Q u i c k S t a r t G u i d e
        desired temperature using the up and down arrow
        keys. Any temperature between 15-60°C can be
        chosen. Each stage can run at a unique temperature.
To choose the desired 
        highlighted and then cycle to the desired value using
        the up and down arrows. Any value between
a
        0.003 and 1.0 
 can be chosen. The step setting
w
        for DVS determines the 
        and consequently the resolution. A step setting of
a
        0.10 
 will result in humidity being controlled to
w
        0.1, 0.2, 0.3 ... up to the final 
        level until either equilibrium is achieved based on
        the trigger value or the timeout value is reached.
        Setting a higher value will make the test faster, but
a
        will decrease the 
 resolution.
w
To choose the trigger %dm/dt value that will indicate
        equilibration, press enter with "Trigger" highlighted
        and then cycle to the desired value using the up and
        down arrows. The trigger can be shut off or be any
        value between 0.001-1%/hr. Choose the number of
        events required to meet the trigger value. The trigger
        value represents an acceptable change in mass per
        change in time to indicate equilibrium for a given step. 
        Setting a higher value will make the test faster, but
        may not result in complete equilibrium. If the trigger
        is turned off, the sample will be held at each step for
        the time indicated by the timeout setting.
To choose to include a timeout setting, press   with
        "Timeout" highlighted and then cycle to the desired
        value using the up and down arrows. Any value
        between 5.0 min to 30 days can be selected. The
        timeout determines the maximum time allotted for
a
        each 
 step NOT the time from starting 
w
a
. The default value is off, which means there is no
w
        timeout value and only the trigger value will
        determine the end of step. If both a trigger value and
        timeout value are set, whichever is reached first will
        determine the end of the step.
To choose to have the isotherm loop, press   with
a
 step, press   with "Step"
w
a
 values for equilibration
w
a
, remaining at each
w
a
 to final
w
          4.  Lightly tap the table with one finger. If the
              readings vary more than 0.2 mg after tapping,
              the table is probably too unstable for accurate
              weight readings during a test and will need to be
              made more rigid or another location found for
              this equipment. Contact your AquaLab
              distributor if the scale readings are unstable after
              placing the unit on a stable structure.
Desiccant Cartridge
Desiccant Receiver
Black Valve
Black Receiving Hole
Figure 2: Desiccant installation
•  Desiccant tubes are installed by orienting the tube so that
  the black colored valve inserts into the black receiving hole
  as shown in Figure 2: Desiccant installation. Before insertion
  make sure that the desiccant tube lid is firmly sealed onto
  the desiccant tube. (See Manual for instructions) Lift the
  tube into the area above the desiccant receiver and then 
  insert the tube into the mating holes while shaking the tube
  slightly. Once in place, the desiccant tube is sealed for flow
  purposes and may move side to side if contacted.
V S A Q u i c k S t a r t G u i d e