4 COMMUNICATION
DS-2
4
Communication
The DS-2 anemometer communicates using SDI-12 protocol. This
chapter discusses the specifics of this method. For more informa-
tion, please visit
http://www.decagon.com/support
for the complete
integrators guide, which gives more detailed explanations and in-
structions.
4.1
SDI-12 Communication
The DS-2 anemometer communicates using the SDI-12 protocol,
a three-wire interface where all sensors are powered (white wire),
grounded (bare wire), and communicate (red wire) on shared wires
(for more info, go to www.sdi-12.org). There are some positive and
negative elements of this protocol. On the positive side, multiple
sensors can be connected to the same voltage supply and commu-
nication port on the data logger. This simplifies wiring because no
multiplexer is necessary. On the negative side, one sensor problem
can bring down the entire array (through a short circuit, etc.). To
avoid this problem, we recommend the user make an independent
junction box with wire harnesses where all sensor wires are connected
to wire lugs so sensors can be disconnected if a problem arises. A
single three-wire bundle can be run from the junction box to the data
logger.
The SDI-12 protocol requires that all sensors have a unique address.
DS-2 anemometers come from the factory with an SDI-12 address
of 0. To add more than one SDI-12 sensor to a system, the sensor
address must be changed to some value other than zero and all sen-
sors on the bus must have unique addresses. Address options include
1..9, A..Z, a..z .
There are two ways to set the SDI-12 sensor address. The best and
easiest is to use Decagon's ProCheck (if the option is not available
on your ProCheck, please upgrade to the latest version of firmware).
SDI-12 addressing can be accessed in the "CONFIG" menu by se-
lecting "SDI-12 Address" and pressing Enter. To change the SDI-12
address, press the up and down arrows until you see the desired
10