Decagon Devices GS3 Manual - Halaman 6

Jelajahi secara online atau unduh pdf Manual untuk Aksesori Decagon Devices GS3. Decagon Devices GS3 12 halaman. Greenhouse sensor

GS3 Integrator Guide
SDI-12 TIMING
SDI-12 communication sends characters at 1200 bits per second. Each character has 1 start bit, 7 data
bits (LSB first), 1 even parity bit, and 1 stop bit (Active low or inverted logic levels):
All SDI-12 commands and response must adhere to the following format on the data line. Both the
command and response are preceded by an address and terminated by a carriage return line feed
combination.
DATA LOGGER
Break
(at least 12ms)
(at least 8.33)
*MAXIMUM TIME
380 mS
Most Commands
780 mS
For a D command after a concurrent measurement
810 mS
For a D command after a concurrent measurement, with CRC
D0
D1
START
Example SDI-12 transmission of the character '1' (0x31)
Command
Marking
ms)
For support contact us via email at
or call us at 1.509.332.5600 between 7AM to 5PM PST
D2
D3
D4
D5
D6
EP
(at least 8.33)
Sensor must respond
within 15 ms
STOP
SENSOR
Response
Marking
ms)
Maximum time*
R34
6