To avoid RCD tripping a test has to be carried out between L and N during
approx. 5s. Immediately afterwards, voltage testing between L and PE
can be carried out without RCD tripping.
5.5. Maintenance
When using voltage testers in compliance with the instruction manual, no
particular maintenance is required. If functional errors occur during
normal operating, our service department will check your instrument
without delay.
5.6. Cleaning
Prior to cleaning, remove voltage test from all measurement circuits. If
the instruments are dirty after daily usage, it is advisable to clean them by
using a damp cloth and a mild household detergent. Never use acid
detergents or dissolvents for cleaning. After cleaning, do not use the
voltage tester for a period of approx. 5 hours.
5.7. Calibration Interval
The voltage testers must be calibrated periodically and checked by our
service department at regular intervals to ensure the specified accuracy of
measurement results. We recommend a calibration interval of one year.