Gamry Interface 1000 Panduan Memulai Cepat - Halaman 2

Jelajahi secara online atau unduh pdf Panduan Memulai Cepat untuk Peralatan Uji Gamry Interface 1000. Gamry Interface 1000 4 halaman. Usb potentiostat

When possible, connect the Chassis Ground on the back of your
potentiostat to a known, good earth ground.
Connect the cell cable to the Calibration side of the UDC4 dummy cell
included with your instrument.
Place the dummy cell inside the Calibration
Shield, close the lid, and connect the black
floating-ground lead of your cell cable to the
Shield's grounding post.
Open Gamry Framework™. Select Experiment > Utilities > Calibrate Instrument
Choose your potentiostat. Select the radio button for calibration type Both, and click the OK button.
(Even though AC calibration for the
Interface™ 1000 is performed at the
Factory, be sure to click the Both radio
button.)
After you confirm several instructional
messages, the calibration proceeds
automatically, and you are notified if the
calibration is successful.