Mounting and commissioning
WARNING
6.7

Zero point adjustment at the remote indicator

14
2.
Press mini button (1) and hold it down. Plug the two-pole battery
plug into the two-pole socket (2) on the PCB.
Fig. 3: PCB
Danger of explosion due to short circuit of lithium battery
Do not short-circuit the lithium battery.
3.
Release the mini button.
The pointer of the direct indication gauge points to zero.
1.
Close the housing of the remote indicator by pushing the hous-
ing parts together.
The remote indicator was switched on when the battery was con-
nected. The display toggles between "Zero" and the current offset of
the complete system. The upper left corner of the display shows the
two arrows
to indicate that the instrument is in calibration mode.
2.
Press the two keys
the offset to the value 0.00.
In this state, zero calibration can be performed any number of
times.
3.
Press the F key in order to terminate zero calibration.
An arrow pointing to the unit "Litres" is displayed at the bottom
of the display.
and
simultaneously in order to correct
Mini button
1
Socket
2
Unimes E