CORNING LSE 6767-HS Buku Petunjuk - Halaman 11

Jelajahi secara online atau unduh pdf Buku Petunjuk untuk Peralatan Laboratorium CORNING LSE 6767-HS. CORNING LSE 6767-HS 16 halaman. High speed microcentrifuge

CORNING LSE 6767-HS Buku Petunjuk
6. Troubleshooting Guide
Please refer to this guide before calling for service.
Centrifuge will not start
Possible reason:
Solution:
Lid lock will not release
Possible reason:
Solution:
Possible reason:
Solution
Possible reason:
Solution:
Possible reason:
Solution:
Centrifuge cannot be started, although power is on
Possible reason:
Solution:
Possible reason:
Solution
bAL: Error Message
Possible Reason:
Solution:
Possible reason:
Solution:
Possible reason:
No power supply
Check that power is being supplied to
the outlet
Check that the power cord is plugged
into both the wall outlet and the back of
the centrifuge
Check that power cord is not damaged
Defective lid lock
Open manually and have unit serviced
No power from PC board
Call for service
Lid lock is jammed
Call for service
Centrifuge is not receiving power
See "Centrifuge will not start"
Lid not closed correctly
Close lid correctly
No speed or time has been selected
Set speed and/or time
Indicates imbalance
Tubes not inserted symmetrically in
rotor holes
Load tube symmetrically
(see Section 3.4 on Loading the rotor)
Sample liquid in tubes not balanced
Make sure the same volume of liquid is
in each tube
Defective or improperly adjusted
balance sensor
8