Calibrating
1. Select pH or conc. mode
2. 1-point calibration
Example
pH
7. 0 00
1
cal
cal buffer/standard reminder
Auto endpoint or
3. Rinse electrode and blot dry
to return to sample measurement, or
continue with step 4 for further cal points
4. 2- (3 pH or 3, 4, 5 concentration) point calibration,
repeat steps 2 and 3 using other buffer/standard
auto or manual endpoint
pH only
95.0 - 105.0%
electrode in
good condition
When you press cal the pH buffer or concentration
standard you selected for cal 1 (Program Menu) is
displayed for 2 seconds, and then the 315 starts
measuring. If you want to use another pH buffer
(choice of 3) or conc. standard (choice of 5) press cal
again. Press cal repeatedly to exit calibration routine.
The decimal point flashes during cal measurement.
When the electrode output has stabilized
the stability indicator appears.
Manual endpoint - press read to endpoint
Auto endpoint - the 315 automatically endpoints.
When a calibration has endpointed you can press
mode to display the absolute mV value and
temperature of the buffer or standard; the display
reverts to the buffer or standard and E
values, or the buffer or standard and slope values
for a 2-(3, 4, 5) point calibration after 2 seconds.
buffer or
standard
2
7. 0 03
seconds
25. 0 °c
cal
ATC
7. 0 00
3. 2
cal
E
0
H
O
2
4. 0 00
97. 2
cal
% slope
90.0 - 94.9%
85.0 - 89.9%
electrode
electrode
needs cleaning
needs
conditioning
(pH only)
0

Measuring Samples

1.
Select pH, mV or concentration mode
2.
Measure sample
Example
pH
1
Auto endpoint or
pH
1
mV
3.
Rinse, blot and store electrode
If you press mode to select pH, mV or concentration
the 315 begins measuring immediately, and you will
not need to press read to start the sample
measurement.
The decimal point flashes during sample
measurement. When the electrode output
has stabilized the stability indicator appears.
pH
Manual endpoint - press read to endpoint
Auto endpoint - the 315 automatically endpoints.
2
6. 8 40
sample
6. 8 40
H
O
2
pH
25. 0 °c
ATC
pH
25. 0 °c
ATC