Avante Volumax Panduan Referensi Cepat - Halaman 3
Jelajahi secara online atau unduh pdf Panduan Referensi Cepat untuk Peralatan Medis Avante Volumax. Avante Volumax 4 halaman. Vos
CANNULA FLOW RATES
CARTRIDGE
High Flow (P04108)
Low Flow (P04108LF)
*VAPCPS is intended to deliver flow rates of 1-20 lpm
ALARM ICON
GENERAL FAULT
(FLASHING)
GENERAL FAULT
(FLASHING)
DISPLAYS DASHES ( - - )
BLOCKED TUBE
(FLASHING)
WATER OUT
(FLASHING)
DISPOSABLE WATER
PATH
BATTERY CHARGING
(FLASHING)
BATTERY
2601 Stanley Gault Pkwy, Suite 101
800-923-8413
502-244-6345
•
CANNULA TYPE
VAPCA, VAPCPAS, VAPCPS*
VAPCPM, VAPCN, VAPCI, VAPCII, VAPCISP,
VAPCPS*
INDICATES
Malfunction of sensor or
control system
O
sensor fault
2
% O
2
High back pressure
No water in disposable
water path. Gas flow
continues without heating
or water circulation
Disposable water path
faulty or not detected.
Unit will not run.
(FLASHING)
The internal battery backup is not fully charged. The unit would not run on battery for the full
rated time in the event of a power failure. No action is necessary.
The unit is running
BATTERY mode. Gas flow
and blending continues
without heat or water
circulation.
(FLASHING)
Louisville, KY 40223 USA
•
FAX: 502-244-0369
•
OPERATIONAL FLOW RATES
5-40 liters per minute (lpm)
1-8 liters per minute (lpm)
CAUSE
Cannot be corrected by user:
Internal component
disconnect patient. Shut off unit,
failure
send for service.
Depleted or defective
Reset by shutting off unit. Replace
O
sensor
O
sensor. Restart unit
2
2
Obstructed or kinked
cannula/delivery tube,
Clear obstruction, check cannula
incorrect cannula
type, re-install DPC
for flow rate or DPC
improperly seated
Disconnect patient. Replace water
Sterile water empty, or
bag or straighten inlet tube. Restart
obstructed inlet tube
unit.
Disposable water path
If disposable water path is present,
defective, not properly
remove and replace to reset
seated or not installed
detector.
AC power is
Reconnect AC power
disconnected
www.avantevet.com
ACTION