1.
Plug out the power connector from the socket.
Remove screw protection caps on the frontside
2.
Loosen 2 screws with philips screwdriver
3.
4. take out the frontplate and electronics board
5. untighten 2 screws from HgAr socket
6. slightly bend the electrical wires out of HgAr
socket
7. take out the HgAr Lamp from Lamp holder
8. replace by new Lamp, make sure the lamp legs
make no short circuit.
9. Slide back electronics board and front plate,
tighten screws and put back screw
Technical Data
Power requirement
Output
Optical power in 600µm fiber
Connector
Internal Voltage
Warm up
Dimensions
Ordering Information
Mercury-Argon Spectral Calibration Source
AvaLight-CAL
AvaLight-CAL-RM Rack mounted version of Mercury-Argon Spectral Calibration Source
Replacement bulb, Mercury-Argon
AvaLight-CAL-B
Power supply 100-240VAC/12VDC, 1.25A for AvaLight-CAL
PS-12V/1.25A
24-08-04
Avantes
HgAr Lamp replacement
HgAr lamp
12VDC supply, 420 mA
Hg and Ar lines from 254 nm to 922 nm
1.6 mW
SMA 905 connector
1200 Volts AC at 30 kHz, 10 mA
1 minute for vapor stabilization
175 x 110 x 44 mm
AvaLight-CAL manual.doc