A.H. Systems Model Active Monopole Antennas

CALIBRATION

Preamplifier Alignment

Preliminary alignment is required on all manufactured active monopoles prior to placing the units in
stock for sale. All preliminary test sheets and ITP will be maintained and a copy stored with the unit
to be placed in stock.
Turn on the equipment and allow 15 minutes for warm up.
Set up the equipment to 50 MHz center frequency, 10 MHz Span/DIV, 2 dB/DIV amplitude, 300
KHz resolution and 80 dBµV reference amplitude on the spectrum analyzer.
Set the output level on the tracking generator to 72 dBµV
Connect the tracking generator to the spectrum analyzer as shown in Figure 3 and the 50Ω load
connected to the antenna signal output port (B).
Save trace as trace A.
Connect the tracking generator cable to the fixture terminated with 50Ω (A) and connect the
spectrum analyzer cable to the signal output port (B) per Figure 3.
Adjust the resistor pots (R
R6 adjusts high-frequency gain
R7 adjusts low-frequency gain
R8 adjusts overall gain
Verify that the low frequency cut off is below specification.
Records providing evidence that the required final inspection and tests are carried out shall be
maintained and shall be available for review by Customer and third party inspectors. Such records
include signed off ITPs, Job cards, Nonconformance reports, Testing reports and Inspection
reports if any.
 A.H. Systems inc. – October 2005
REV A
, R
and R
Figure 1) to match the following curve (Figure 2).
6
7
8
Figure 1
Figure 2
10