Aqua Creek Products Mighty 600 Manual - Halaman 10
Jelajahi secara online atau unduh pdf Manual untuk Bantuan Mobilitas Aqua Creek Products Mighty 600. Aqua Creek Products Mighty 600 14 halaman. Pool lift
WARNING! DO NOT ATTEMPT TO DIRECTLY PROBE THE BATTERY CONNECTOR TO TEST VOLTAGE.
THIS MAY CAUSE PERMANENT DAMAGE TO THE BATTERY! USE THE SUPPLIED BATTERY TEST CABLE
ONLY FOR CHECKING THE BATTERY VOLTAGE!
The Aqua Creek 24V battery has a cable and connector for connecting to the control box or to the
battery charger. The battery connector is not suitable to check the battery voltage, as it is very hard
to probe the connector without shorting-out the battery, which will cause damage. For this reason, a
separate battery test cable is provided. To safely check the battery voltage, use a voltmeter or multi-
meter with standard test probes, and insert the probes into the test cable as shown below.
10
Copyright © 2019 Aqua Creek Products All Rights Reserved
Battery Test Cable Instructions
CHECK VOLTAGE
24V BATTERY
BATTERY
CONNECTOR
MALE
CONNECTOR
BATTERY
TEST
CABLE
INSERT TEST
PROBES TO
Revised 4/23/19