AB Sciex API 5000 LC/MS/MS Manual Perangkat Keras - Halaman 11

Jelajahi secara online atau unduh pdf Manual Perangkat Keras untuk Peralatan Laboratorium AB Sciex API 5000 LC/MS/MS. AB Sciex API 5000 LC/MS/MS 16 halaman.

AB Sciex API 5000 LC/MS/MS Manual Perangkat Keras
Figure 1-8 Source exhaust system
WARNING! Hazardous Materials: Take all necessary precautions to ensure the safe
disposal of the source exhaust gas. Follow all applicable local regulations.
A filtered nitrogen or zero air gas supply (oil free) is delivered to the source
exhaust pump at pressures indicated on the gas panel at the back of the
instrument. The previous figure shows the exhaust supply connection points for
the API 4000™ and API 5000™ systems.

Vacuum Chamber

The vacuum chamber houses the mass filter rail, including the ion optics, the
quadrupole rod sets, the collision cell, and the ion detector.

Shutting Down and Turning on the System

Use the following procedures if you need to shut down or turn on the system.
To shut down the system
Document Number: D1000092206 B
RF FEEDBACK
RF FEEDBACK
B
A
B
A
Front Bulkhead Assembly
GAS 1 / GAS 2
MAX105 PSIG
1. In the Analyst
software, complete or stop any ongoing scans.
®
2. Turn off the sample flow to the instrument.
Ion Source
Pressure Switch
Exhaust Supply
BACKING PUMP
CURTAIN GAS SUPPLY
MAX 60 PSIG
EXHAUST SUPPLY
EXHAUST WASTE OUT
MIN 55 PSIG
MAX 60 PSIG
Exhaust Waste Out
Gas Panel
Hardware Guide
Venturi Pump
11