4
CALIBRATION
A calibration process must be performed before
the indicator is ready to indicate tension. Select
an appropriate calibration weight. The weight
determines the value of web tension that will
produce full output of the 1100TV. For example: A
10 lb weight will result in full output of 100 lbs of
tension if the 10% calibration is performed. If a
calibration of 25% is used, a calibration weight of
25 lbs will result in a full scale output of 100 lbs
of tension.
1. Setup
In the lower right corner of the main display, tap
the green Menu tab. Now tap the Calibration
Menu tab. You may be prompted to press the
Request Access tab, if this occurs enter 6 1 5
0 on the number pad. After gaining access, tap
Calibration Menu again to proceed.
Next, tap Calibrate. Select the units of
measure for display (Fig. 5). The choices are
lb, oz, g, Kg or N. After selection, press OK.
You will now need to select a measurement
range or scale (Fig. 6). Swipe left or right to
view all available options. Press OK.
Next, select the percentage of your tension
range for calibration (Fig. 7).
Note: The 10% calibration weight ratio is
recommended for high tension applications, or
any time it may be a challenge to handle a
heavy calibration weight. The 25%
calibration ratio is recommended for maximum
accuracy and is especially useful in calibrating
for low tension applications.
Make your selection and press OK.
Review your selections (Fig. 8) and press OK
to accept them all or BACK to return to prior
screens and makes changes.
Figure 5 - SELECTION OF TENSION UNITS
Figure 6 - SELECTION OF TENSION RANGE
Figure 7 - SELECTION OF CALIBRATION WEIGHT
Figure 8 - REVIEW OF CALIBRATION SETTINGS