Some basic checks are detailed below. If in doubt contact service support on 0870 066 0809.
Bath Does not move
 Check Power.
 Check Spur Fused to ensure there is power is to the unit.
 Ensure the RCCB is working and has power to it. To test the RCCB push the test button; the
RCCB should interrupt power and the switch should trip. Push the RESET button; power
should be restored. If the RCCB fails to operate in this manner there is a problem with the
electric supply and the service department should be contacted.
If mains power is interrupted the bath will only work on battery backup. The bath will continue to operate
but as the batteries become low the bath will emit an audible beep when the handset is used.
Once the warning Beep starts, lower the bath and do not use until the fault is rectified. Once power is
restored allow the batteries to recharge before operation. (Usually 3 hours)
If the bath fails to operate, reset the system:
Depress both 'Bath' buttons on the handset at the same time – 10 long beeps will sound and the system will
reset. It is worthwhile checking that the handset cord and jack are still firmly located into the control box
Bath judders.
 Usually occurs if the bath has not been fixed to the floor correctly or the bath is not level.
Bath noisy
 Check to ensure there is nothing touching the bath while it moves
 Ensure the pipe-work and cables are free of the panels and not rubbing against the framework
 Check panels are free to move and are not touching each other.
Bath Leaking
 Look for source of leak – typically caught flexi-hose on feed or waste. This can usually be remedied
by your own maintenance contact.
Bath is uneven
(one end higher than the other)
 This may occur after a power failure. Use the DOWN button on the handset. Lower the bath to its
lowest position. This will automatically reset the bath so that both sides rise together.
Bath Beeping
Mains power has failed to the unit – See section titled 'Bath does not move'
Abacus Pisces 069/Info.P 04/14v1a

TROUBLE SHOOTING