3.2.1.

Serial port configuration

Select the serial port (COM port) on your computer that is connected to the
instrument by the 'special drop-down menu (formito the USB cable
supplied is a virtual COM port).
From the menu of the instrument "8. PC DATA "enable data connection,
so that the instrument is in state" CONNECTING ... "
3.2.2.

Exporting saved data from instrument memory

Select whether to export to a text file or an
excel file (you can select both)
Then clicking on the "RECEIVING DATA FROM MEMORY" the data on the device is exported to
the previously selected file
NOTE: The instrument must be in the "CONNECTION ...."
Example of the exported file in excel:
3.2.3.

Setting up company data

Type in the fields with white background data
to be loaded.
Then clicking on the "DATA ENTRY
COMPANY" 's header is loaded on the
instrument,
at
this
automatically the received data in order to
assess their correctness pa.
NOTE: The instrument must be in the "CONNECTION ...."