Filter Replacement
Cold evaporation filter Replacement
The cold evaporation filter should be replaced
every eight months, or whenever indicated
by the yellow 1 illuminated on the Filter
Replacement Indicator on the Control Panel.
1. Remove the top cover from the back of the
Sterilizer.
2. Grasp the handle of the HOCl generator
and turn it to the right to release it from
the base.
3. Remove the HOCl generator.
NOTE: Handle the Cold evaporation filter
by the end caps to avoid touching the Cold
evaporation filter and preventing possible
contamination of the filter element.
4. Unscrew the cold evaporation filter from
the generator and discard.
5. Remove the replacement cold evaporation
filter from packaging and screw it into the
inner bottom of the HOCl generator.
6. Replace the HOCl generator in the
Sterilizer and turn the handle to the left to
lock it into place on the base.
7. Turn on the Sterilizer. After replacing the
filter, press and hold the Filter Replacement
Button for 3 seconds to reset the filter life
countdown.
Note that If you replace the filter before
the accumulated life time is reached, the
filter life timer cannot be reset.
4-in-1 HEPA Filter Replacement
The filter should be replaced every 18 months
or whenever indicated by the red 2 illuminated
on the Filter Replacement Indicator on the
Control Panel. However this interval is intended
as a guideline only. In environments with a high
concentration of contaminants such as dust,
11
pet dander, or cigarette/cigar smoke you may
need to replace the filter more frequently.
NOTE: The HEPA filter is not washable. DO
NOT immerse it in water.
1. Make sure the Sterilizer is OFF and
unplugged.
NOTE: If you have difficulty removing the
bottom cover, remove the top cover prior
to removing bottom cover.
2. Detach and remove the magnetic bottom
cover from the unit.
3. Remove using the pull tabs and replace
the replacement 4-in-1 HEPA filter into the
Sterilizer main body.
NOTE: Ensure that the pull tabs are facing
outward to facilitate filter removal later.
4. Inspect the screen on the bottom cover
and if necessary clean any excess dust or
dirt from the screen using a hand held
vacuum.
5. Replace the bottom cover by placing the
bottom cover into the slots on the unit.
Gently push the top of the bottom cover
into place.
6. Turn on the Sterilizer. After replacing
the filter, press and hold the Filter
Replacement Button for 3 seconds to
reset the filter life countdown.
Only use genuine D&HLifelabs
CleVaFilter&AquaSoFilter with your AquaSolvo
Eco Ambient Air Sterilizer. Use of any other filters
may compromise the proper operation of your
unit and potentially void your warranty.
•4-in-1 HEPA Filter: Part Number DH02FT-1
•Cold Evaporation Filter: Part Number
AS04FT-2
To purchase replacement filters, remote control, or
accessories visit www.DHLifelabs.com
Troubleshooting
Symptom
Sterilizer does not operate.
Sterilizer is starting to make noise and
the air output is low.
Air coming from Sterilizer has a bad
odor.
Air quality does not improve as the
Sterilizer operates.
Water added to Sterilizer but Water
Level indicator still flashing.
Salt Level indicator displayed even after
salt has been added.
Salt Level indicator flashing after salt
has been added.
The appliance is very loud.
Probable Cause/Solution
Power cord not securely plugged into outlet. Make sure plug
is securely inserted.
No AC power. Check fuse or circuit breaker. Replace fuse
with a time delay type or reset the circuit breaker.
4-in-1 HEPA filter may be dirty or obstructed. Replace filter.
HOCl generator is dirty or has contaminated water in it.
Discard any water that is in the generator and thoroughly
clean it, then refill with clean fresh tap water and non-iodized
salt.
NOTE: The first time you use the Sterilizer, you might
perceive a plastic or paint smell. This is normal, but it
should disappear quickly. If the smell persists or if you sense
a burning smell, please disconnect the unit and contact
Customer Assistance.
Windows or doors are open. Constant introduction of dirty
air from outside the room can reduce the effectiveness of the
Sterilizer. Close all exterior doors and windows.
Check that HOCl generator is properly seated in the
Sterilizer.
It takes some time to dissolve the salt and monitor the actual
concentration. If necessary discard salt water solution and
mix a new batch.
Make sure amount of salt added is correct. If necessary
discard salt water solution and mix a new batch.
Ensure that there are no obstacles near the appliance and
that it has a minimum of 12 inches (30 cm) of clearance
around it for optimal operation.
12