2- Installation

2.1 Accessories Kit

The following accessories are included in the DTN-410's kit:
1 Main supply chord
3 Fuses for type F, 3 amps
3 Thermal printer paper rolls
1 Plastic shaft for thermal paper roll
1 User's manual
1 Pre-focused 6 v – 10 w halogen lamp
1 Hexagonal Wrench
1 #1 cross-recess screwdriver
1 Sample aspiration capillary tube
1 Peristaltic pump tube (kit x 3)
1 Drain hose
1 Set of sample filters
1 Cleaning solution #1
1 Soaking solution #2
1 Plastic semi micro cuvettes kit x 5
Check if the packing is in good condition before opening, and inform about any irregularity.
After unpacking, inspect if all items are in good condition. Please report non-conformances
to manufacturer or authorized dealer immediately.

2.2 Main supply voltage

The analyzer is a Class I instrument that requires 100 to 240 VAC 50/60 Hz.
This equipment requires a single-phase line with protective ground connection.
The protective ground connection must fulfill regulatory
requirements according to the voltage available at the place
where the equipment is being connected. Check the protective
ground connection before installation.
Required fuse is type F (fast) for 3 amps.

2.3 Cautions

1) Place the instrument in well-ventilated environments, free from dust and corrosive
fumes.
2) Do not install the equipment near to heaters, muffles, ovens, etc. Avoid places
subject to drastic changes in temperature.
3) Do not use tables subject to intense vibrations. Vibrations caused by unbalanced
centrifuges may affect the optic system.
DIALAB PHOTOMETER DTN-410K USER MANUAL