Treat P (CPAP Mode only): Indicates CPAP Treatment Pressure. Pressure Range
from 4 to 20 cm H2O in 0.5 cm H2O increment.
Init P (CPAP Mode only): The starting pressure if a Ramp time is set. Setting range
CLINIC MENU
is 4 cm H2O to Treatment P (Prescription Pressure). Default setting is 4 cm H2O.
Auto
CPAP
S Mode
S Mode
Treat P
epap
epap
Init P
ipap
ipap
Ramp
Ramp
Ramp
Init ipap
Max ipap
Rise Time
init epap
Rise Time
isense
isense
esense
Ti Min
esense
Ti Max
Ti Min
Ti Max
Unit
Use Time
Date
Save
EPAP (S and Auto S Mode):
Expiratory Positive Airway Pressure.
Setting range is 4-25 cm H2O.
Cannot go above IPAP Pressure.
Default setting is 6 cm H2O.
IPAP (S and Auto S Mode):
Inspiratory Positive Airway Pressure.
Setting range is 4-25 cm H2O..
Default setting is 8 cm H2O.
Ramp P: the starting pressure
after blower is turned on when
RAMP Time is set at anything above
0 minutes. The setting range is
4cmH2O to Treat P or Init EPAP
Init EPAP (S and Auto S Mode):
When a Ramp time is set, this will
be the initial pressure to begin ramp
pressure rise.
Max IPAP (Auto S Mode only):
This setting is the maximum IPAP
Pressure. Default setting is 25 cm
H2O.
Ti Max (S and Auto S Mode):
Maximum Inspiration Time. It refers
to the maximum duration of IPAP in
each respiratory cycle. Setting range
is 0.3-3.0 Seconds. Adjustment step is
0.1 S. Default setting is 2.0 S.
Ti Min (S and Auto S Mode): Minimum Inspiration Time. It refers to the
minimum duration of IPAP in each respiratory cycle. Setting range is 0.1-Ti Max.
Adjustment step is 0.1 S. Default setting is 0.5 S.
Rise Time (S and Auto S mode): is the time it takes for the device to transition
from EPAP to IPAP once inspiration has triggered. It is adjustable by the provider
from 0 to 3, default setting is 2.
I Sense (S and Auto S mode): This Setting can adjust the trigger sensitivity when
patient's breath switches into inspiratory phase. It can be adjusted from 1 (less
sensitive) to 8 (most sensitive).Default setting is 6
E Sense (S and Auto S mode): This Setting adjusts the trigger sensitivity when
a patient's breath switches into expiration phase. It can be adjusted from 1 (less
sensitive) to 8 (most sensitive). Default setting is 7.