How to Wash and Disinfect
1. Wash hands
2. Prepare detergent (soapy water) and disinfectant solution wearing personal protective
equipment (PPE): gown or impermeable apron as a minimum, eye protection and
surgical mask for bleach solution
3. Wash first in soapy water, to remove blood and respiratory secretions. In clean sink or
bucket brush the equipment under water to prevent splash and ensure all visible soiling
is removed, rinse with water that has been boiled but allowed to cool to tepid
4. Let dry
5. Wash next in diluted bleach or disinfectant. Bleach needs one hour of soaking. Soak all
items together, do not keep putting one in and taking one out. Once used bleach
should not be re-used or kept in storage, discard after use.
6. Rinse with water that has been boiled and cooled to tepid (rinse also inside, for
example, using a sterile syringe), let it drip dry over the sink, do not leave it coiled on
the sink.
7. Remove all protective gear (apron, gloves, mask) and wash hands thoroughly
8. Check that there is no pooled water in the circuit. Store the circuit and bottle in a clean
plastic bag (labeled and dated). Store in the dry and clean area (separate from a soiled
equipment area)
Routine Cleaning and Maintenance of the CPAP-Concentrator
Each week the oxygen concentrator will require approximately 30 minutes of attention.
Concentrators have a large particulate filter over the air inlet opening (usually at the back of
freestanding or portable models). This filter stops dust and other airborne particles from entering
the unit. The filter should be removed and cleaned in warm soapy water, completely dried with an
absorbent towel and replaced. Have a spare dry filter to replace with so there are minimal
interruptions to the concentrator function.
The exterior of the oxygen concentrator should be cleaned with a mild disinfecting cleaning agent
or a diluted solution of bleach (usually 5.25% sodium hypochlorite). A solution in the range of
1:100 to 1:10 of bleach to water can be used effectively. Allow the solution to remain on the
surface for 10 minutes and then rinse off and dry.
Accessories and spares
All accessories used with the Baby CPAP must:
 Be oxygen compatible,
 Be biocompatible,
 Comply with the general requirements of the 93/42/EEC European Directive
A full list of available spares is available by contacting Diamedica – [email protected]