ABB Endura AZ40 Petunjuk Pengoperasian - Halaman 5

Jelajahi secara online atau unduh pdf Petunjuk Pengoperasian untuk Peralatan Industri ABB Endura AZ40. ABB Endura AZ40 16 halaman. Oxygen and carbon monoxide equivalent (coe) analyzer
Juga untuk ABB Endura AZ40: Manual (16 halaman), Panduan Petunjuk Pengoperasian (20 halaman)

ABB Endura AZ40 Petunjuk Pengoperasian
5.3 Disconnecting the sensor electrical power and
signal cables at the process
DANGER – Serious damage to health / risk to life
Allow sufficient time for the sensor assembly to cool
before performing this procedure.
DANGER – Serious damage to health / risk to life
The transmitter must be isolated from mains power
supplies before performing this procedure.
Referring to Fig. 5.3:
1. Use a medium flat-bladed screwdriver to unscrew 4
(captive) cover screws A and remove cover B from
sensor assembly C.
2. Disconnect mains cable D from terminal block E
3. Disconnect signal cable F from terminal block G.
4. Disconnect thermocouple cables H from terminal block
I.
5. If optional blowback is fitted, disconnect cable J from
terminal block K.
Fig. 5.3 Disconnecting sensor cables
5.4 Removing the sensor assembly from the process
DANGER – Serious damage to health / risk to life
Allow sufficient time for the sensor assembly to cool
before performing this procedure.
Referring to Fig. 5.4:
1. Use a 10 in. adjustable spanner (wrench) to remove 4 nuts,
washers and lockwashers A securing sensor assembly
B to mounting flange C. Set items aside for re-use.
2. Carefully remove the sensor assembly and the attached
probe (including filter assembly) from the process.
3. Temporarily cover process opening until the sensor
assembly is ready to be re-installed.
Fig. 5.4 Removing the sensor assembly from the process
Endura AZ40 | Sensor assembly | Orifice and seals kit | INS/ANAINST/007-EN Rev. B
5