10.6 Replace fuse
1). Switch off the power.
2). Push inward with a flat head
screw driver, then unscrew the fuse
holder (
counter clockwise).
4).Replace the fuse with a fuse
of the proper electrical
requirements (see pg.2).
.
11. Transport and Storage
11
.1 Switch off the sterilizer before transportation or storage. Pull out the plug
and let the machine cool down.
11.2 Drain both the clean and the used water tank.
11.3 Conditions for transportation and storage:
Temperature: -20 C /~ +55C
Relative humidity: 85%
Atmospheric pressure: 50kPa~106kPa
1 5
3). Pull out the fuse holder by
hand.
5). Place the fuse holder back in
place and use a flat head screw
driver to fully push it in. Then
tighten (clockwise).
Instructions manual
6.6 After the samples are loaded, close
and lock the door by turning the door
handle clockwise until it stops.
Caution: You must turn the door handle to the maximum position,
otherwise the machine will alarm and an error message will
be displayed during the cycle.
6.7 Starting the sterilization program.
Press the START button, the machine will begin
the sterilization cycle and complete all phases
(fill, sterilize, exhaust and dry) automatically.
The cycle will complete in 12-65 minutes.
(See
)
Appendix 2
Caution: When you press the "Start" button and the door handle has not
been turned to the maximum position, "
on the Time display window. The cycle will not begin
until the door has been completely closed and the "Start"
button is pressed again.
6.8 Sterilization cycle completion
After a cycle is completed the time display will show "Ed"
Once the pressure has reached 0 PSI on the display the door can be
opened and samples can be removed.
Always use the tray removal tool to load or unload the tray in
order to avoid touching the hot surface of the trays.
Instructions manual
will flash
do"
8