Dickson KT6P2 Memulai - Halaman 3
Jelajahi secara online atau unduh pdf Memulai untuk Perekam Suara Dickson KT6P2. Dickson KT6P2 7 halaman. Remote sensing temperature recorder
Juga untuk Dickson KT6P2: Panduan Memulai Cepat (9 halaman)
Pen 1 alarm minimum
Pen 1 alarm maximum
3. In order to set pen alarm minimum and maximum, pressing Adjust-Up will increase the alarm value,
and pressing Adjust-Down will decrease the alarm value. There is acceleration if the Adjust-Up button
is held down.
Repetitively pressing the Pen Home button will scroll through the three options until the Alarm
button is pressed to exit from alarm adjust. Each press of the Pen Home or Alarm button will store the
new settings. The unit will take one minute to exit Alarm Set Mode once the Alarm button is pressed.
4. If the alarm is triggered, the LED will show as solid red and the alarm will sound. Press the Alarm
button to silence the audible alarm.
Alarm Relays (KT6P5)
The SPST 24V 500mA relay contacts are normally open and will close on alarm conditions when the
alarms are enabled. Relay contacts are always functional when the alarm is enabled. The relay will close
only during minimum and maximum alarm conditions. Relays will not operate under battery only power
conditions
DIP SWITCH SETUP
To setup the KT6 recorder for your specific application, you might need to change some of the Dip
Switches. The Dip Switches are located on the back of the unit. (Figure 3)
Use a pen or small screw driver to flip the switches. Remember to install the correct chart to match
the corresponding Dip Switch setting.
Recording Time: The KT6 recorder has two different recording time options
7 Day
24 Hour
F/C: Record in degrees F or C
F
C
Temperature Range: Select the temperature ranges that matches the chart paper
0 to +100 F/C
-50 to 0 F/C
-100 to 0 F
-50 to 50 F/C
0 to 50 F/C
Dickson KT6P Manual V3
1.800.561.8187
#2 Down
#2 Up
#3 Down
#3 Up
#4 Down
#5 Down
#6 Down
#4 Down
#5 Down
#6 Up
#4 Down
#5 Up
#6 Up
#4 Up
#5 Down
#6 Down
#4 Down
#5 Up
3
www.
.com