Alpha Moisture Systems DS5000 Buku Petunjuk - Halaman 15
Jelajahi secara online atau unduh pdf Buku Petunjuk untuk Monitor Alpha Moisture Systems DS5000. Alpha Moisture Systems DS5000 19 halaman. Dewpoint monitor
InFo-
9.10.1
rmation Details
LCD Display
Software version.
Transmitter Serial Number - upper digits.
Transmitter Serial Number – lower digits.
Transmitter Lower Range.
Transmitter Upper Range.
ACAL value.
ACAL point
Offset Implementation Point.
Offset Value.
If both offset values are the same, or 0, then there is no current offset
Current Engineering Units.
Alarm 1 trigger (Hi or Lo).
Alarm 1 value.
Alarm 2 trigger (Hi or Lo).
Alarm 2 value.
Bell on / off.
Pressure Units.
LP Pressure value.
SP Pressure value.
4-20mA Lo range.
4-20mA Hi range.
Voltage Lo range.
Voltage Hi range.
10 Normal Operation of the DS5000
In normal operation, the DS5000 will display the current value as received via RS485 from
the connected DT45 transmitter. The value is updated once per second and is displayed in the
currently selected engineering units, which is indicated by the LED.
If a pressure has been entered into the DS5000, the LED will alternate between the symbol
for the currently selected engineering units and the '≡' icon. This indicates that the m oisture
values displayed on the LCD have been pressure corrected.
The Alarm LED's (AL 1 & AL 2) will light whenever an alarm condition occurs and only
turn off when the alarm condition clears. The DS5000 has an internal buzzer, which also
sounds whenever an alarm condition occurs. The buzzer can be muted by pressing the '↵'
key. The buzzer automatically shuts off whenever an alarm condition clears.
1188 Instruction Man DS5000.doc
Example
V0.18
Sn20
-207
-80.0
20.0
0.1
4.7
-52.0
-40
C
Hi
20.0
Lo
-20.0
On
BArg
0.0
0.0
-80.0
20.0
-80.0
20.0
LED Display
Blank
S
S
L
U
A
r
P
o
.
E
1
1
2
2
b
»
4
Ü
v
Ü;
Page 12 of 16