Olympus DP71 Petunjuk Manual - Halaman 23
Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Kamera Digital Olympus DP71. Olympus DP71 32 halaman. Microscope digital camera
Juga untuk Olympus DP71: Petunjuk Manual (32 halaman)
Trigger Output
· The trigger output can be used to control a commercially available shutter.
· The positive and negative logics of the trigger output can be switched with the DP Controller.
· The trigger output is interlocked with the shutter release/close operations of the DP Controller.
· The external trigger signal is output by starting still image recording when the shutter is closed.
· The time lag from the trigger output to the start of exposure can be set between 0 and 2 sec. on the DP Controller.
· The trigger output is a TTL compatible signal.
V
: 2.4 V (min.). V
: 0.4 V (max.).
OH
OL
Trigger output
Exposure
Readout &
transfer to
PC
Trigger Output Timing Chart in Still Image Recording (Positive logic)
Exposure time lag
DP71
20