Olympus BH-2 BHT Pembongkaran, Pembersihan, dan Pemasangan Kembali Lengkap - Halaman 8

Jelajahi secara online atau unduh pdf Pembongkaran, Pembersihan, dan Pemasangan Kembali Lengkap untuk Mikroskop Olympus BH-2 BHT. Olympus BH-2 BHT 23 halaman. Sliding focus block
Juga untuk Olympus BH-2 BHT: Manual (24 halaman)

Olympus BH-2 BHT Pembongkaran, Pembersihan, dan Pemasangan Kembali Lengkap
Separate the Arm from the Base 
Separate  the  arm  from  the  base  of  the  stand  per  the 
following procedure. 
Remove the Lamphouse 
Remove  the  lamphouse  from  the  base  by  grasping  it 
and pulling it straight back (see Figure 3).  Do not touch 
the  halogen  lamp  with  your  fingers,  as  oils  from  your 
skin  may  cause  premature  failure  of  the  bulb.    If  the 
bulb  is  accidentally  touched,  clean  it  with  isopropyl 
alcohol. 
Figure 3 – Remove the lamphouse from the base 
Remove the Electrical Base 
Carefully  turn  the  stand  upside  down.    Use  a  4mm  (or 
5/32") Allen wrench or driver to remove the four M5x12 
hex,  socket‐head  cap  screws  and  the  associated 
washers from the bottom of the stand (see Figure 4).  
Figure 4 – Bottom view of the BHT/BHTU electrical base 
Hold the electrical base in place and set the stand in its 
normal upright position on a flat surface, being careful 
not to let the electrical base shift as you do so.  Lift the 
stand from the electrical base by grasping the arm and 
lifting  straight  up  until  the  bottom  of  the  stand  well 
clears the electrical base.  If the stand contains a light‐
Complete Teardown, Cleaning, and Reassembly of the Olympus BH‐2 Sliding Focus Block  
 
preset  control/switch  (see 
disengage  the  preset  control/switch  from  the  notch  in 
the side of the base
electrical base.  Set the electrical base aside. 
Figure 5 – Light pre‐set control and pre‐set switch 
Remove the Arm from the Base 
Lay the stand on its back.  Use a 5mm Allen wrench or 
driver to remove the four M6x25 hex, socket‐head cap 
screws securing the arm to the base (see Figure 6) and 
separate the arm from the base (see Figure 7).   
 
Figure 6 – Remove screws securing the arm to the base 
 
Figure 7 – Separate the arm from the base 
                                                            
2
 Or unplug the two electrical connectors from the printed circuit board. 
Figure 
5),  be  sure  to 
2
 before lifting the stand clear of the 
Revision 2 
 
 
 
 
Page 8 of 23