Use  a  suitable  solvent  (e.g.,  acetone)  to  remove  any 
grease from the four long guide rails (see Figure 19) and 
from the side grooves in the arm casting (see Figure 20).  
Figure 19 – Clean grease from the guide rails 
Figure 20 – Clean grease from the side grooves in the arm 
Remove and Reinstall the Bottom Dust Cover 
The  function  of  the  bottom  dust  cover,  which  is  not 
present on all BH‐2 stands, is to prevent dust and debris 
from  entering  the  opening  at  the  bottom  of  the  arm 
(where  the  arm  attaches  to  the  base  casting)  and 
fouling the mechanism of the sliding focus block. 
Although this dust cover, if present, does not need to be 
removed to perform the maintenance described in this 
document,  its  removal  may  be  desirable  to  allow  for  a 
more‐thorough  cleaning  of  any  dust  and  debris 
accumulation  in  the  area  of  the  dust  cover,  and  the 
removal procedure is included in this document for the 
sake of completeness.   
Use a suitable JIS screwdriver to remove the two M2x4 
screws securing the dust cover onto the bottom of the 
arm  casting  (see 
Figure 
cover  (see 
Figure 
22)  and  clean  any  dust  and  debris 
from the cover and from the associated area of the arm 
casting.  Reposition the dust cover in place and secure it 
to the arm casting with the two M2x4 screws. 
Complete Teardown, Cleaning, and Reassembly of the Olympus BH‐2 Sliding Focus Block  
21).    Lift  and  remove  the  dust 
 
Figure 21 – Remove screws securing bottom dust cover 
 
Figure 22 – Remove the bottom dust cover 
 
Remove the Guide Rails from the Focus Block 
Remove the two short guide rails from the side groove 
in the focus block (see Figure 23). 
Figure 23 – Remove two guide rails from the focus block 
Use  a  suitable  solvent  (e.g.,  acetone)  to  remove  any 
grease  from  two  short  guide  rails  (see 
from the side groove in the focus block (see Figure 24).  
 
Figure 
19)  and 
Revision 2 
Page 11 of 23