Olympus BX-FLA Petunjuk Manual
3. Preparing
the Power Supply Unit
0 When the transmitted ligrht source is used simultaneously,
the built-in power supply of the BX microscope is not usable.
The TH3 power supply unit is required in this case.
When only the vertical illumination is used, the built-in power supply of the BX50 microscope can be used. (Connect
the LH connector to the microscope connector.)
\
I
Fig. 24
4. Centering
the Halogen
Lamp
.
. .
Fig. 25
I. Connect the lamp housing cord connector to the output connector @
on the power supply unit. (Fig. 24)
2. Connect one end of power cord to the AC power cord receptacle 0,
and the other end to an AC outlet.
Circuit Breaker Reset
(Fig. 24)
The circuit breakers @ will activate in case of brightness adjustment
circuit malfunction (short circuit, etc.) or in case of overcurrent. When
they activate, the center portion of the breaker will pop up and the
power will be cut off.
To reset the breakers, press in the center portion. If a breaker activates
again, unplug the power cord from the AC outlet and contact your
Olympus representative.
0 Before proceeding to the following
procedure, remove the analyzer,
polarizer and filters from the light path.
1. Engage the 1 OX objective.
2. Place a highly reflective specimen (e.g. mirror) on the stage, and bring
the image into approximate
focus.
3. Remove one of the eyepieces, and make adjustment so that the filament
image in microscope sleeve will be positioned at center.
Focusing of filament image:
Turn the bulb focusing knob 0.
Vertical adjustment:
Turn the bulb up/down adjustment
knob 0.
Horizontal adjustment:
Turn the bulb left/right adjustment
knob 0.
4. Return the removed
analyzer, polarizer and filters to their original
positions.
0
40