Adjusting the X- and Y-Axis Knob Tension

1. Hold the X-axis knob @ and slide up the Y-axis knob 2 up to expose
2
the adjustment knobs.
2. Turning the X-axis adjustment knob 3 or Y-axis adjustment knob |
clockwise (in the direction of the arrow) increases the tension and
3
counterclockwise decreases it.
CAUTION
1
} After long hours of use, the stage guide may be deviated and the stage
travel range may be decreased. However, this is not malfunction and can
be corrected easily as described below.
[Treatment]
Horizontal direction: Hold the specimen holder and move the stage guide
Vertical direction:
Stage Feed Knob Rubber Caps (Optional)
}When the X- and Y-axis knobs are fitted with the rubber caps, the knobs
can be adjusted without slipping and fine adjustment is possible by
holding the knobs with a very light force. The knob rubber caps also
reduce fatigue after long hours of operation.
The U-SHGT thick type (thickness 5 mm) and U-SHG thin type (thickness
2 mm) knob rubbers are available.
To attach the knob rubbers:
First fit the larger knob rubber to the Y-axis (upper) knob from below it,
then fit the smaller knob rubber to the X-axis (lower) knob from below it.
3

Rotating the Stage

1. Using the Allen screwdriver, slightly loosen the stage clamping screw @.
2. The stage can be rotated both clockwise and counterclockwise by the
stage clamping screw.
CAUTION
} The angle of rotation varies depending on the positions of the X- and
Y-axis knobs.
If the tension is adjusted to tight, a creaking sound may be heard
during stage travel, and the stage stopping accuracy may be
imperiled.
to the left and right so that it hits the stoppers.
Hold the upper stage and move it to the front and rear
so that it hits the stoppers.
A click may be heard and felt during rotation. However, this is
due to the construction of the substage and does not indicate
a malfunction.
Clockwise
Right hand knobs
230°
Left hand knobs
20°
(Fig. 22)
(Fig. 23)
Angle of Rotation
Counterclockwise
20°
230°