Olympus BX46 Petunjuk Manual - Halaman 26

Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Mikroskop Olympus BX46. Olympus BX46 44 halaman. Clinical microscope

Olympus BX46 Petunjuk Manual
1

5-5 Condenser

For how to adjust the centering of the condenser, see page 12.
1
Compatibility of Objectives and Condensers
Objective Magnification
2X/4X
10X/20X/40X
} The 1.25X to 2X objectives can also be used but the surroundings of the field of view may become dark.
22
2
Fig. 32
Applicable by swing the top lens out. Used as the aperture
Applicable by engaging the top lens
in the light path.
6

Using Eyepieces Incorporating a Micrometer

} When the eyepieces in use incorporate a micrometer, the accuracy of
the left-right focusing adjustment (diopter adjustment) can be improved
further.
1. Looking into the right eyepiece with your right eye, turn the top of the
eyepiece @ so that the micrometer in the field of view looks sharpest
(Fig. 32).
2. Looking into the right eyepiece, rotate the coarse and fine adjustment
knobs to bring the specimen and micrometer (cross lines, etc.) into
focus.
3. Looking into the left eyepiece with your left eye, rotate the diopter
adjustment ring 2 to bring the specimen into focus.
BX45 Condenser
Field iris diaphragm
Aperture iris diaphragm
Opened
iris diaphragm
Field iris diaphragm
Aperture iris diaphragm
(Fig. 32)