Problem
e) Coarse adjustment will not go all
the way down.
f ) Objective makes contact with speci-
men before focus is obtained.
4. Observation Tube
a) Field of view of one eye does not
match that of the other.
5. Stage
a) Image shifts when you touch
stage.
b) Specimen stops midway on the
X-axis traverse.
c) Horizontal and/or X- and Y-axis kn-
obs are too tight or too loose.
d) Stroke has reduced.
30
Cause
Condenser holder is too low.
Specimen is mounted upside down.
Interpupillary distance is incorrect.
Incorrect diopter adjustment.
Different eyepieces are used on left
and right.
Your view is not accustomed to micro-
scope observation.
Stage is not properly mounted.
Specimen is not correctly positioned.
Tension of horizontal and/or X- and
Y-axis knobs is too high or too low.
Stage guide is deviated.
Remedy
Raise condenser holder.
Mount specimen correctly.
Adjust interpupillary distance.
Adjust diopter.
Change on eyepiece to match the
other so that both sides are the same
type.
Upon looking into eyepieces, try
looking at overall field before concen-
trating on specimen range. You may
also find it helpful to look up and into
distance for a moment before looking
back into microscope.
Clamp stage.
Place specimen correctly.
Adjust tension.
Use treatment method described on
page 20.
Page
12
11
11,22
21
19
20
20