Replacement of Cubes
2-2 Replacing the Spectral Cube
To improve the efficiency of fluorescence detection, the fluorescence waveform
separating dichroic mirrors and barrier filters (2 channels) of the spectral cube inside the
external photo-multiplier (FV10-OPD) can be replaced according to the excitation
wavelength to be used.
2-2-1 Removing the spectral cube
1.
2.
II. PREPARATION For OPERATION
II. 2 - 4
Page
/ Replacing the Spectral Cube
Using an Allen screwdriver, loosen the two cover clamping screws (provided with
slip-off prevention mechanisms)
remove the cover
.
Loosen the cube cover clamping screws
step 1, and remove the cube cover by holding the cover knob
on the left side panel of the FV10-OPD and
inside the cover in the same way as in
.