Olympus IX71 Brosur & Spesifikasi - Halaman 4

Jelajahi secara online atau unduh pdf Brosur & Spesifikasi untuk Mikroskop Olympus IX71. Olympus IX71 16 halaman. Microscope z-axis motor drive
Juga untuk Olympus IX71: Panduan Instalasi (11 halaman)

Olympus IX71 Brosur & Spesifikasi
Two-tier multi-port design ensures input/output flexibility.
Improved near-infrared transmission
With the introduction of the new UIS2 optical system, the IX2 series
offers improved IR transmittance for the side port, back port and
bottom port, providing versatile, high-performance response to
future research demands.
IX2 Side port: Transmittance improved by new coating
100
90
80
70
60
50
40
30
20
10
0
300
400
500
600
700
800
900
1000
Wavelength (nm)
Current IX2 left side port
New IX2 left side port
100
Upper Tier Lightpath Selection (optional)
Located between objective and tube lenses so a parallel
90
pencil of rays can be obtained or introduced. Primary
image can be gained by adding a tube lens. Inserting
optical components such as a dichromatic mirror does
80
not produce a double image. (The alternative of the right
side port)
70
60
Lower Tier Lightpath Selection (included)
Located below the tube lens inside the frame, this tier
allows primary image access to either the left side port
50
or lower back port.
40
30
102 mm from left side port
mounting position to primary image.
20
Bottom Lightpath Selection
10
A direct primary image can be obtained without any
reflections and any optical components.
0
300
400
5
Lower back port
IR compatible
Left side port
IR compatible
Bottom port
500
600
700
Two-tier optical design is also near-IR compatible
The input/output of a parallel pencil of rays and the multiple port
structure for gaining the primary image are designed internally in the
form of tiers. To maximize the possible wavelength width, the
optical path branching of each tier is also compatible with the near-
infrared spectrum. Even when more than one port is being used
simultaneously, there is no change in the stage height; as a result,
rigidity and illumination performance remain constant.
Binocular port
Right side port
IR compatible
Lower back port
Enables the attachment of a
C mount camera. Only 1X
magnification is available.
Bottom port /
IX2-TVR (T-mount)
Primary image access is also
available at the microscope bottom.
800
900
1000
Optical Port system
IX2 Two-tier optical path
Optical path as seen from
Optical path as seen from
the left side of
the front of
the microscope
the microscope
Objective
Mirror unit
Tier-1:
Upper optical path
Right side port
selection
Tube lens
Tier-2:
Left side port
Lower optical path
Lower back port
selection
Bottom optical
path selection
Right side port / IX2-RSPC-2
The right side port unit (IX2-RSPC-2:
option, F.N.: 16) comes with a tube
lens and accepts a C-mount CCD
camera.
Left side port
This port offers a high quality
primary image, located in 102 mm
distance from the microscope
frame for maximum flexibility in
mounting filter wheels or any kinds
of camera adapters.
Dual port camera adapter /
U-DPCAD* (C-mount,
left side port)
This unique dual port adapter
enables the provision of two
primary images suitable for live cell
imaging.
* optional unit
6