6. The light intensity adjustment does not interlock at the zooming.
6. The light intensity adjustment does not interlock at the zooming.
NO
Assemble zoom volume ass'y in
zoom gear.
B. TROUBLESHOOTING
B. TROUBLESHOOTING
Refer to "Illuminator LED does not light up"
and "Brightness adjustment knob does not
operate" in troubleshooting.
NO
When the zoom turns, the axis of
volume resistance is rotated through
gear.
YES
Zoom volume ass'y has damage.
or light intensity board has damage.
(Change them.)
B-6
B-6
The light intensity adjustment is
NO
possible by brightness adjustment knob.
YES
The light intensity adjustment does
not interlock in all of zoom area.
YES
The illuminator should be the
maximum by brightness adjustment
knob already.