Olympus MX61 Panduan Perakitan - Halaman 34

Jelajahi secara online atau unduh pdf Panduan Perakitan untuk Mikroskop Olympus MX61. Olympus MX61 49 halaman. 200mm/300mm compatible semiconductor/fpd inspection microscopes

Olympus MX61 Panduan Perakitan
6

Command Details

Log-in

Functions

1. Switches between remote and local modes of the MX61/61L.
2. Checks the current mode (local or remote) of the MX61/61L.

Special notes

1. When the microscope is in local mode, it ignores commands except for "1LOG
2. When the microscope is in local mode, it cannot receive the present commands during operation of the aperture iris
diaphragm or revolving nosepiece.
3. When the microscope is set to remote mode by this request command, all of the front panel switches, the external control
of revolving nosepiece (page 38), I/F signal input and the aperture iris diaphragm control input according to the DF/BF
sensor are disabled except for the light intensity control and pilot lamp, and the microscope becomes exclusively
controllable with these commands from the host. The illumination in the front panel switches and hand switch buttons
are extinguished in remote mode.
4. When the microscope is set to local mode by this instruction command, all of the front panel switches and hand switch
buttons are enabled and their illumination is lit up.
Command descriptions
Command
1LOG
IN
1LOG
OUT
1LOG?
)
"
" represents a space.
*
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.
31
Type
Description
Request
Local
remote mode
switching.
Request
Remote
local mode
switching.
Query
Check of the
current local/re-
mote mode.
Response
Result
1LOG
+
Successful
1LOG
!,E01511
Error
1LOG
!,E01120
Error
1LOG
+
Error
1LOG
!,E01120
Error
1LOG
IN
Successful
1LOG
OUT
Successful
1LOG
!,E01150
Error
1LOG 1LOG?
IN" and the query commands.
Description
The microscope is switched to
remote mode.
It cannot be switched to remote
mode because one of motorized
units is operating.
A string other than "IN" and "OUT"
was sent after "1LOG
".
It is switched to local mode.
A string other than "IN" and "OUT"
is sent after "1LOG
".
The microscope is in remote
mode.
The microscope is in local mode.
A string is sent after "1LOG?
".