Olympus SZ61 Petunjuk Manual - Halaman 23
Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Mikroskop Olympus SZ61. Olympus SZ61 28 halaman. Zoom stereo microscopes
Juga untuk Olympus SZ61: Petunjuk Manual (28 halaman)
2
External View and Nomenclature
Clamping screw mount holes
Stage
Load capacity: 1 kg
Clamping knob
3
Installation
1. Using the provided screw and Allen wrench, attach the SZ2-FO up/down moving stage into the screw hole on the stage
plate mount hole of an applicable base or that on the stage plate (SZ-STL).
The SZ2-FO can also be installed so that the focus adjustment knobs are located on the near side of the operator. This
orientation is obligatory when the IHE holder for the LSGA oblique illumination system is used.
2. Loosen the clamping knob on the stage, attach the stage plate, SZH-SG or SZH-SC, and tighten the clamping knob
again.
4
Operation
@
Push up.
Fig. 17
Clamp.
Fig. 18
Removing the stage plate (Fig. 17)
To remove the stage plate, loosen the clamping knob @ and push up
the stage plate from below it.
Adjusting the focus adjustment knob tension (Fig. 18)
}This operation is intended to facilitate the rotation of the knobs while
To increase
preventing spontaneous drop of the up/down moving stage. It is recom-
mended to set the knob tension to a slightly higher level than the point
where spontaneous drop occurs.
a. Hold the left and right focus adjustment knobs @ with both hands, fix
@
the left knob and rotate the right knob. The rotation tension of the
To decrease
knobs increases or decreases according to the direction in which the
right knob is rotated.
b. If the rotation tension is increased too much, accurate focusing is not
possible and the mechanism may be damaged.
SZ2
Focus adjustment knobs
Stroke: Approx. 21 mm
Rotation tension adjustable.
series
20