Installation
# Before mounting, remove dirt and dust from the mount surfaces and handle carefully so as not to damage them.
Stage plate
Friction surfaces
Stage plate mount
hole
3
Operation
CAUTION
#
Do not touch the friction surfaces on the cup stage and seat with a hand. If the friction surfaces are contaminated,
wash with a neutral detergent before use.
# If a load of more than 20 grams is applied to the edge of the cup stage, it may move spontaneously.
# When a tall specimen is placed and the cup stage is tilted, the specimen may go out of focus. In this case, adjust
the focus again.
23
Specimen holder
Specimen holder
mount hole
Cup stage
Seat
Fig. 20
Fig. 21
Fit the cup stage seat into the stage plate mount
1
hole of an applicable base.
Place the cup stage on the seat.
2
Before placing, wipe the friction surfaces on the
cup stage and seat with a clean cloth.
Mount the stage plate.
3
Mount the specimen holder.
4
}Clean the friction surfaces periodically.
Applicable base
Place a specimen on the stage plate, hold the cup stage by the edge,
and tilt the cup stage slowly. (Fig. 20)
}If the specimen slips on the stage plate, hold the specimen with the
provided specimen holder.
}To fix a container such as a petri dish, insert the provided tube into the
specimen holder to fix the container by pinching. (Fig. 21)