9-5 BX Stage Adapter Type 2 SZX-STAD2

}This adapter is for installing the U-SIC4R2/SIC4L2 large stage* on the SZX2-STL/SZX-STL large stand**. When this
stage adapter is used, the SZH-P400 auxiliary pillar should be used to cover the height of the stage adapter (about 125
mm). In addition, when a low-magnification objective with a long working distance (DFPL0.5X-4) is used, the SZH-P600
auxiliary pillar should be used (always together with the drop prevention collar).
* The U-SVL or U-SVR BS stage for the BX series microscopes can also be mounted but the operability deteriorates. The
U-SVLB and U-SVRB cannot be used due to their long stage knobs.
** A SZX2/SZX series illuminator stand can also be mounted, but its built-in transmitted illumination cannot be used.
1
Assembly
Frosted filter
BX stage adapter type 2
SZX-STAD2
Mirror knob
Large base
SZX2-STL
# For simplified transmitted light observation, place the mirror knob on the front and use a frosted filter.
³
²
@
Large stage U-SIC4R2/SIC4L2
Allen wrench
(provided with the base)
Clamping screw
Mounting threaded hole
SZX10