Attaching the Lamp Housing
Installing the Mercury Burner
1. Loosen the socket clamping screw @ using the Allen screwdriver.
2. Hold the upper section of the lamp housing and pull it upward to remove
the socket section.
# To prevent malfunction, do not hold the lamp housing by the centering
knobs ².
3. Place the socket section upside down as shown in Fig. 12.
}The lamp housing is equipped with the holder for transportation in the
factory shipment condition or with an old burner when the burner is
replaced. Remove the holder or old burner by loosening the two burner
holding screws ³.
4. Attach the + (positive) pole of a specified mercury burner | to the fixed
mount on the upper side, then the - (negative) pole to the mount on the
|
lower side.
# Be sure to use the USH-103OL (OLYMPUS) or HBO103W/2 (OSRAM).
To prevent burner cracking due to glass distortion caused by stain, be
careful not to stain the burner with fingerprints or dirt. If it is contaminated,
clean by wiping gently with a piece of gauze moistened with absolute
alcohol.
5. Attach the socket section with burner to the original position and tighten
the socket clamping screw @.
# Align the external edges of the lamp housing with those on the
socket section, and push the lamp housing straight downward.
}This value assumes light cycles composed of 2 hours of lighting
and 30 minutes of extinction. Do not turn it on and off at a shorter
cycle than the above, for this will shorten the service life of the burner.
After replacing the burner, reset the hour counter to "0.0" as
outlined above.
Installing the Lamp Housing
1. Using the Allen screwdriver, loosen the two clamping screws @ on the
lamp housing mount hole.
2. Fir the lamp housing all the way into the hole as shown in Fig. 13.
3. Tighten the clamping screws @ using the Allen screwdriver.
# The lamp housing and its surroundings become extremely hot. When
installing the microscope system, reserve ample spaces around the
lamp housing, particularly above it. Also be sure not to install the
lamp housing obliquely.
SZX16 REFLECTED FLUORESCENCE SYSTEM
Burner Service Life
USH-103OL: 300 hours
(Figs. 11 to 13)
12