Olympus SZX9 Petunjuk Manual - Halaman 23

Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Mikroskop Olympus SZX9. Olympus SZX9 34 halaman. Research stereomicroscope system

Olympus SZX9 Petunjuk Manual

OPERATION OF OTHER ATTACHMENTS

9-1 Eyepoint Adjuster SZX-EPA

1
External View
2
Assembly
19
Observation tube clamping screw
@
²
Fig. 27
1. Using the Allen screwdriver provided with the SZX microscope body,
remove the observation tube @.
2. Mount the eyepoint adjuster ² to the place where the observation tube
has been mounted.
3. Mount the observation tube (removed in step 1 above) onto the eyepoint
adjuster.
This allows you to increase the eyepoint height by 40 mm. Up to two
eyepoint adjusters can be mounted provided that no other intermediate
attachment is used.
(Fig. 27)