Olympus VS120 Panduan Pengguna - Halaman 11

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Mikroskop Olympus VS120. Olympus VS120 44 halaman. Virtual slide system
Juga untuk Olympus VS120: Fitur (2 halaman)

Olympus VS120 Panduan Pengguna
7
Confirming Settings
+
Acquiring Images /
Lines for the number of the slides to be scanned
are shown.
Click the lines you want to work on.
Change the objective lens as needed.
Click
to edit the scan area.
Use the
mark in the Edit screen to edit the scan
area, and then press the [Enter] key on the keyboard.
Click
to edit the focus map.
Confirm the
(focus map) in the Edit screen.
Delete the focus map by right clicking on it and selecting
[Delete] from the menu.
After editing the focus map, press the [Enter] key on the
keyboard.
The method for adjusting the focus can be set using
[Focus Logic]. If [Semi-automatic focusing] is selected,
focus can be adjusted manually after the focus is
automatically set.
For more information on Focus Logic, refer to Section
"1-4 Setting Focus Logic" (page 29) or the online help.
Click
to select the next slide.
Repeat steps from
to
which you need to set slide glass information.
VS120-L100
(Expert Batch)
for all the slides for
7
Page