ESG-300 Operation Settings
Basic Procedures
Procedure
Snare Polypectomy:
Sessile polyp
Snare Polypectomy:
Pedunculated polyp
Endoscopic Mucosal
Resection (EMR)
Hemostasis
Sphincterotomy / Pre-Cut
Endoscopic Submucosal Dissection (ESD)
Procedure Steps
Marking
Pre-Cut / Incision
Submucosal Dissection
Hemostasis
Peroral Endoscopic Myotomy (POEM)
Procedure Steps
Incision
Dissection / Tunneling
Myotomy
Hemostasis
The above settings are examples based upon previous general clinical usage of the device. Please note that clinical staff are solely responsible for selection of an
appropriate mode and power level, dependent upon the instrument being used and the condition of the tissue being treated. Due to the heterogeneity of instruments
and clinical conditions, Olympus cannot accept liability for clinical risks arising from the use of these example settings.
Instrument or technique
Close the snare
slowly and gently
Close the snare
slowly and gently
Close the snare
slowly and gently
Coagrasper
Sphincterotome / Needle
Knife
Instrument or technique
Knife (apply gentle pressure)
Knife
Knife
Closed Knife
Instrument or technique
Knife
Knife
Knife
Coagrasper
Mode
PulseCut Slow
ForcedCoag
PulseCut Slow
SoftCoag
PulseCut Fast
Mode
ForcedCoag
SoftCoag
PulseCut Fast
BlendCut
PowerCoag
PowerCoag
Mode
PulseCut Fast
SprayCoag
PulseCut Fast
SoftCoag
Power
Effect
120
2
20
4
120
2
50
3
120
2
Power
Effect
20
2
50
3
120
2
30
2
30
2
30
2
Power
Effect
120
2
40
2
120
2
50
3