Olympus K-404 Panduan Referensi Cepat - Halaman 2
Jelajahi secara online atau unduh pdf Panduan Referensi Cepat untuk Peralatan Medis Olympus K-404. Olympus K-404 2 halaman.
<Preparation of ET Stopper>
Insert the EndoTherapy device into the guide sheath's
accessory port so that the device tip protrudes slightly
from the guide sheath tip. Attach the ET stopper to
the EndoTherapy device by inserting the EndoTherapy
device to the end of the slit.
Note:
Make sure the EndoTherapy device
reaches the end of the slit on the
stopper.
Tips
Adjust the position of the ET stopper so that the tip of each device protrudes from the guide sheath tip as
shown below:
Guiding device
3 Attachment of Stopper to the Guide Sheath
Attach the ET stopper
1
to the guide sheath with
the ultrasonic probe or
EndoTherapy device
inside.
Notes · Do not insert the guide sheath into the endoscope when it is not combined with any EndoTherapy accessory or
ultrasonic probe. Doing so could result in bending or breaking the guide sheath.
· When inserting the guide sheath, hold it close to the biopsy valve and keep it as straight as possible, then insert it
gently into the endoscope.
Caution
If the endoscope's insertion section near the control
section is bended, it may be difficult to advance the
EndoTherapy device into the guide sheath smoothly.
Make sure to keep the insertion section near the
control section of the endoscope as straight as
possible.
As medical knowledge is constantly growing, technical modifications or changes of the product design, product specifications, accessories and service offerings may be required.
ET stopper
Biopsy forceps
Postbox 10 49 08, 20034 Hamburg, Germany
Wendenstrasse 14-18, 20097 Hamburg, Germany
Phone: +49 40 23773-0, Fax: +49 40 233765
www.olympus-europa.com
Cytology brush
Insert the guide sheath
2
with stopper into the
endoscope. When
it reaches the target
position, move the
stopper attached to
the guide sheath until it
makes contact with the
biopsy valve.
Straight
Aspiration needle
Bended
N8602374-122021