4-1
Main Unit Specifications
4-1-1
Applicable Endoscopes
OES Scope
EVIS Scope
4-1-2
Pump Used
Electromagnetic vibrator pump
4-1-3
Air Feeding Performance
At the scope socket:
4
2.84 x 10
Pa or greater at 0 ml/min;
4
1.76 x 10
Pa or greater at 2000 ml/min;
Note: Measured at the rated power supply voltage and a room temperature of 20 to 25 ºC
4-1-4
Control Method
By turning the power switch ON/OFF.
4-1-5
Hook Installation
Can be hung down by connecting the hanger to the hook installed in a wall.
(The hook is screwed into a wall and fixed.)
4-1-6
External Dimensions
Width 85 x Height 157 x Depth 170 mm
4-1-7
Weight
1.8 kg
4-1-8
Power Supply
Voltage:
Frequency:
Current:
Voltage fluctuation rate:within ±10%
4-2
Ambient Environment
4-2-1
Ambient Temperature
10 to 40ºC
4-2-2
Relative Humidifty
30 to 85%
ISSUE6
4. Specifications
100 V, 120 V, 220 to 240 V AC
Both 50/60 Hz
0.4A (100 V, 120 V), 0.2A (220 V to 240 V)
1-3
MU-1
Product Outline