Olympus ORBEYE OME-V200 Panduan Referensi Cepat
4 Place an object having an easy-to judge vertical
direction view so that it is focused on the center
of the image displayed on the observation
monitor. Observe the observation monitor
images without wearing the 3D eyeglasses and
confirm that there is no positioning error in the
up-down direction between the left and right
images.
5 Observe the images by wearing the 3D
eyeglasses, close the left and right eyes
alternately, and confirm that there is no difference in color or brightness between the left and right images.
6 While observing the 3D images, contact the observed object with forceps or a similar tool and confirm that
there is no peculiarity in the feeling of depth.
7 Confirm that the data display is displayed on the observation monitor.
8 Observe your hand and confirm that the image is displayed
in normal colors without noise.
2.8 Inspection of the focus/zoom functions
1 Tilt the scope switches A and B on the microscope
section all the way upward or downward and, by
observing the observation monitor, confirm that the
focus and zoom functions are activated correctly.
2 If the zoom + of the adaptive AF is set, confirm that
the auto focus is activated and focus is adjusted after
completion of the zoom + (UP) operation.
2.9 Inspection of the view field movement
1 Tilt the joystick on the foot switch all the
way. Observe the observation monitor and
confirm that the field moves to direction in
which the joystick is tilted.
2 If the view field movement of the adaptive
AF is set, confirm that the auto focus is
activated after completion of the view field
movement operation.
Left image
Right image
Images without up-down
positioning error
Observation monitor
Observation monitor
Observation monitor
– 5 –
Left image
Observation
monitor
Images with up-down
positioning error
Observation
Data display
monitor
Joystick
Foot switch
Right image
Scope
switch A
Scope
switch B