Olympus PSD-10 Petunjuk Manual - Halaman 15
Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Peralatan Medis Olympus PSD-10. Olympus PSD-10 28 halaman. Electrosurgical unit
5-3 Connection of Footswitch
Red mark
5-4 Connection of S-P Cord
Red mark
P-cord connector
Terminal
S-cord
connector
Patient-
S-cord connector
plate
mount
Disposable
patient plate
Lever
Clamp
Terminal
Connection of disposable patient plate
Connect the footswitch to the FOOT SW. connector on the PSD-10, with
the red positioning dot on the plug facing upward. Fully tighten the clamp-
ing ring on the plug by turning clockwise.
Connect the S-P cord to the PATIENT connector on the PSD-10, with the
red dot facing upward. Push the plug until it is locked in place.
Connect the P-cord connector to the Patient-plate. Fully tighten the clamping
ring.
Connect the S-cord to the endoscope (S-cord connector mounted on the
light guide connector), which has been plugged in the output socket of the
light source. Fully tighten the clamping ring.
The alarm will be activated when the power switch is turned on, if above
connections (Steps
through
In case a disposable patient plate (Scotchplate
Lift the lock lever of the special S-P cord (MB-583, optional) and insert the
terminal of the patient plate into the port. Press the lever back down to secure
plate.
Non-Controlled Document
are defective.
®
#1149, 3M) is being used: