2-1 Preparation for optical adjustment (CKX31)
Adjustment item
1. Revolving axis
(right sleeve)
2. Left/right optical axis
(left sleeve)
3. Absolute optical axis
(right sleeve)
4. Exit pupil center
5. Parfocality
6. Stage squareness
D. REPAIR PROCEDURE
KN0048 (with adapter-1)
KN0029
FT36
KN0003
Eyepiece(Jig)
1) Standard eyepiece
(KN0048; Adapter-1)
1) Standard eyepiece
(KN0048; Adapter-1)
1) Standard eyepiece
(KN0048; Adapter-1)
1) Centering telescope
(KN0029)
1) Standard eyepiece
(KN0048; Adapter-1)
2) Focusing telescope
(FT-36 or U-FT)
D-4
KN0041
PlanC10X
SKN0003
Objective(Jig)
2) PlanC10X
2) PlanC10X
2) Standard objective
(KN0041)
2) PlanC10X
3) Standard objective
(KN0041)
1) Stage for checking
stage tilt alignment
(SKN0003)
Jigs and tools
3) Test plate :
concentric circles
(KN0003) or
cross micrometer
(OB-M1/100SQ)
3) Test plate :
concentric circles
(KN0003) or
cross micrometer
(OB-M1/100SQ)
2) Thickness gauge
0.3mm
(OT1949)